| Brand Name: | Kenida |
| Model Number: | KDR-200 |
| MOQ: | 1SET |
| Price: | 60000usd/set |
| Payment Terms: | any payments are received |
| Supply Ability: | 1000sets/month |
Lightweight Digital Radiography Equipment , 19″ Medical Color LCD Display
High frequency
With Mobile photography bed
Varian flat panel detector
Main configuration
| No | Part name | Type | quantity |
| high frequency high voltage generator | YJG30R-A | 1set | |
| operating console | YJO-Ⅳ | 1set | |
X-ray tube | H1083 | 1set | |
beam limiting device | M-38 | 1set | |
| flat panel detector | 1800R | 1set | |
| SLC | YJT-Ⅰ | 1set | |
| DR rack | YJDR-Ⅰ | 1set | |
| Mobile photography bed | YJS-Ⅲ | 1set | |
| Acquisition workstation (computer) | color screen | 1set | |
GRID | 103line | 1set | |
| high voltage cable | 75kV | 1pair |
Main technical parameters
| No | Technical parameters |
| 1 | x-ray generator (YJG30R-A) |
| 1.1 | Supply voltage:Three-phase380V±38V |
| *1.2 | Maximum output power:30kW |
| *1.3 | Operating frequency≥100kHz |
| 1.4 | Photography tube voltage:Can continuous ,Adjust the step length of 1 kv, the adjusting range is 40 kv ~ 150 kv |
| *1.5 | Photography tube current:Can step adjustment, adjustment range,10mA~200mA |
| *1.6 | Load time:adjustable range,2ms~6300ms |
| 1.7 | Current-time:Can step adjustment, adjustment range,0.1mAs~500mAs |
| 2 | Operating console(YJO-Ⅳ) |
| 2.1 | DC12V |
| 3 | X-ray tube (H1083) |
| 3.1 | Maximum working voltage: 125kV |
| *3.2 | Bi-focus:0.6mm(small focal point)/1.2mm(big focal point) |
| 3.3 | Anode input power:15KW(small focal point);30KW(big focal point) |
| 3.4 | The anode optical Angle:12° |
| *3.5 | The anode rotation speed:3000rpm |
| 3.6 | Anode heat content:150KHU |
| 4. | beam limiting device(M-38) |
| 4.1 | Lead door control mode:manual operation |
| 4.2 | Power supply: 24VAC 7A |
| 4.3 | Inherent filtration:1.2mmAL |
| 4.4 | Light flash:30s |
| 5 | Flat panel detector(1800R) |
| *5.1 | Fluorescent material: Direct growth cesium iodide |
| *5.2 | Scanning the area size:17×17inch |
| 5.3 | Spatial resolution:≥3.0 lp/mm |
| 5.4 | Analog-digital conversion:16bit |
| 6 | Pixel region:431.8 × 431.8 mm2 |
| 6.1 | Pixels array:2816 × 2816 |
| 6.2 | Pixel size:154μm |
| 6.3 | Image read time:1~1.5s |
| 6.4 | Time of exposure:Shortest:1 ms; longest:10 s |
| 7 | DR rack |
| 7.1 | Stand column hight:2240mm |
| 7.2 | UC arm to move up and down along the column:700mm~1700mm |
| 7.3 | UC arm along the column rotation range:0°~105° |
| 7.4 | Move up and down detector range:700mm~1700mm control mode:Electric braking force:≥100N |
| 7.5 | X ray tube component movement range:0~800mm±50mm Locking way:power-driven;braking force:≥100N |
| 7.6 | SID distance1000mm~1800mm |
| 8 | Mobile photography bed(YJS-Ⅲ) |
| 8.1 | Panel size bed:length×width× height=2020mm×690mm×680mm |
| 8.2 | Operating mode:manual operation |
| 8.3 | Load-bearing:After under 135 kg weight should be able to work properly |
| Others: | |
| 9 | workstation |
| 9.1 | CPU: ≥2.8GHz |
| 9.2 | RAM:≥2GB |
| 9.3 | Disk size:≥500GB |
| 9.4 | Display:≥19″Medical color LCD display |
| 10 | Software function(DRimage processing software) |
| 10.1 | High voltage generator can be equipped with a variety of X-ray tube, the tube of all of the power curve, cooling curve, automatic control by software |
| 10.2 | Has self-check protection function of fault alarm and shutdown in a timely manner |
| 10.3 | Dicom3.0 interfaces for image transmission, query, print, etc |
| 10.4 | With Magellan3 image processing software |